Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.
Showing 41–43 of 43 recalls for “Mislabeling”
Encore Medical, Lp
One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually 25mm.
Terumo Cardiovascular Systems Corporation
During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc