Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.
Showing 541–560 of 1,000 recalls
Siemens Healthcare Diagnostics, Inc.
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Maquet Cardiovascular, LLC
There is a probable mislabeling of product. A Hemashield Platinum Woven Double Velour Bifurcated Vasuclar Graft, SN 1259735369, 24 mm x 12 mm, REF Number M00202166241P0 is likely packaged inside of the carton instead of a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1260004700; 18 mm x 9 mm; REF Number M002020851890 or a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1259882880; 18 mm x 9 mm; Ref Number M002020851890.