Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.
Showing 4781–4800 of 39,430 recalls
Philips Medical Systems Technologies Ltd.
A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.
Shimadzu Corp. Analytical & Measuring Instruments Div.
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.