  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 4701–4720 of 39,430 recalls

Class IIOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4483; use in Peritoneal Dialysis

Baxter Healthcare Corporation

There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.

October 23, 2024 · DeviceView details →
Class IIOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4482; use in Peritoneal Dialysis

Baxter Healthcare Corporation

There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.

October 23, 2024 · DeviceView details →
Class IIOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4484; use in Peritoneal Dialysis

Baxter Healthcare Corporation

There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.

October 23, 2024 · DeviceView details →
Class IIOngoing

URINE KIT W/BD CUP, SKU DYLAB1001A; Component No. 503581

MEDLINE INDUSTRIES, LP - Northfield

Specimen container included in kits is labeled as sterile, but has been identified to be non-sterile. Long-term consequences may include unnecessary treatment and/or prolonged hospitalization.

October 23, 2024 · DeviceView details →
Class IIOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number T5C4482; use in Peritoneal Dialysis

Baxter Healthcare Corporation

There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.

October 23, 2024 · DeviceView details →
Class IIOngoing

URINALYSIS KIT 4, SKU DYLAB1021; Component No. 503581

MEDLINE INDUSTRIES, LP - Northfield

Specimen container included in kits is labeled as sterile, but has been identified to be non-sterile. Long-term consequences may include unnecessary treatment and/or prolonged hospitalization.

October 23, 2024 · DeviceView details →
Class IIOngoing

Ivalon CELLULOSE SURGICAL SPEAR- Intended To Be Used During Ophthalmic Surgical Procedures For Fluid Management Model Number: Q604237

Carwild Corporation

Potential for package seal to be out of specification and the sterile barrier cannot be guaranteed for the life of the product.

October 22, 2024 · DeviceView details →
Class IIOngoing

Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site. Model Number: Q604206

Carwild Corporation

Potential for package seal to be out of specification and the sterile barrier cannot be guaranteed for the life of the product.

October 22, 2024 · DeviceView details →
Class IIOngoing

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Micro-X Ltd.

Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

October 22, 2024 · DeviceView details →
Class IIOngoing

Ivalon Anatomical Nasal Packing with Airway Tube and String, 8.0 cm x 1.5cmx 3.0 cm- Intended To Be Used During Ophthalmic Surgical Procedures For Fluid Management Model Number: Q770410

Carwild Corporation

Potential for package seal to be out of specification and the sterile barrier cannot be guaranteed for the life of the product.

October 22, 2024 · DeviceView details →
Class IOngoing

Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 1197

ICU Medical, Inc.

ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that these non-OEM batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. OSI Batteries and their customers are distributing these non-OEM CSB batteries without authorization from ICU Medical.

October 22, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482S; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4482; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →
Class IIOngoing

Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC

Karl Storz Endoscopy

Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.

October 21, 2024 · DeviceView details →
Class IIOngoing

Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635

Siemens Healthcare Diagnostics, Inc.

Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.

October 21, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4483; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4482E; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →
Class IIOngoing

Medline Iris Scissors, Single-Use, Curved, Standard 4.5 , Model Nos. DYNJ04049 (cases), DYNJ04049H (eaches)

MEDLINE INDUSTRIES, LP - Northfield

Tip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility.

October 21, 2024 · DeviceView details →
Class IOngoing

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp EX Short, Part Number T5C4484; use in Peritoneal Dialysis

Baxter Healthcare Corporation

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

October 21, 2024 · DeviceView details →