  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 3381–3400 of 39,430 recalls

Class IOngoing

Z-800F Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

Zyno Medical LLC

Unreleased software versions were installed on distributed devices without verification or validation.

May 7, 2025 · DeviceView details →
Class IIOngoing

DxI 9000 Access Immunoassay Analyzer C11137

Beckman Coulter, Inc.

Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.

May 7, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Powered Laser Surgical Instrument Product Name: Soltive SuperPulsed Laser Fibers - Single Use Model/Catalog Numbers: TFL-FBX150BS, TFL-FBX200BS, TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TLF-FBX940S, TFL-FBX150BS-JA (Japan Only), TFL-FBX550S-JA (Japan Only), TFL-FBX940S-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150S-JA (Japan Only), TFL-FBX200S-JA (Japan Only), TFL-FBX365S-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.

Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

May 6, 2025 · DeviceView details →
Class IIOngoing

PhenoMATRIX

Copan WASP

AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

May 6, 2025 · DeviceView details →
Class IIOngoing

Affinity 4 Birthing Bed, Product Codes P3700E100429, P3700E100432, P3700E100433, P3700E100434, P3700E100435, P3700E100436, P3700E100437, P3700E100438, P3700E100439, P3700E100441, P3700E100442, P3700E100443, P3700E100444, P3700E100445, P3700E100446, P3700E100448, P3700E100449, P3700E100450, P3700E100453, P3700E100455, P3700E100456, P3700E100457, P3700E100459, P3700E100463, P3700E100468, P3700E100472, P3700E100476, and P3700E100477

Baxter Healthcare Corporation

There is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.

May 6, 2025 · DeviceView details →
Class IIOngoing

HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323

Hardy Diagnostics

Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae.

May 6, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/Catalog Number: TFL-PLS, TFL-PLS-JA (Japan only); Software Version: N/A Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: N/A

Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

May 6, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog Number: TFL-FBX150R, TFL-FBX200R, TFL-FBX365R, TFL-FBX550R, TFL-FBX940R, TFL-FBX940R-JA (Japan Only), TFL-FBX365R-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150R-JA (Japan Only), TFL-FBX200R-JA (Japan Only), TFL-FBX550R-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.

Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

May 6, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number: TFL-SLS, TFL-SLS-JA (Japan only); Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No

Olympus Corporation of the Americas

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

May 6, 2025 · DeviceView details →
Class IIOngoing

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Olympus Corporation of the Americas

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

May 5, 2025 · DeviceView details →
Class IIOngoing

Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system

Dexcom, Inc.

A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.

May 5, 2025 · DeviceView details →
Class IIOngoing

SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

ASP Global, LLC. dba Anatomy Supply Partners, LLC.

All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.

May 5, 2025 · DeviceView details →
Class IIOngoing

Access Testosterone assay, Catalog Number 33560

Beckman Coulter, Inc.

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

May 5, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Canturio¿ Smart Extension Product Name: CSE Implant Model/Catalog Number: 43-5570-030-14 Product Description: Tibial stem extension for use with Zimmer Persona¿ Personalized Knee System; 14mm diameter x 30mm height. Implantable device with integrated electronics for passive kinematic data collection. Single-use, sterile. Packaged in a sterile blister and outer shelf pack. Sold individually. Component: Yes. The Canturio¿ Smart Extension (CSE) is a tibial implant component intended for use with the Zimmer Persona¿ Personalized Knee System.

Canary Medical, Inc.

Due to labeling error. Product incorrectly labeled.

May 2, 2025 · DeviceView details →
Class IIOngoing

Catalog No. 8888160119; Umbilical Vessel Catheter Insertion Tray with 2.5 and 5 French UVCs

Cardinal Health 200, LLC

Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.

May 2, 2025 · DeviceView details →
Class IIOngoing

ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

Healthmark Industries Co., Inc.

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

May 2, 2025 · DeviceView details →
Class IIOngoing

Catalog No. 8888160325; Umbilical Vessel Catheter 2.5 French single lumen

Cardinal Health 200, LLC

Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.

May 2, 2025 · DeviceView details →
Class IIOngoing

BD BBL CultureSwab EZ Collection and Transport System, REF 220144, in vitro Diagnostic, SterileR, Rx Only

Copan Italia

Swabs for specimen collection may be prone to breakage during product usage.

May 2, 2025 · DeviceView details →
Class IIOngoing

Catalog No. 8888160333; Umbilical Vessel Catheter 3.5 French single lumen

Cardinal Health 200, LLC

Packaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates.

May 2, 2025 · DeviceView details →
Class IIOngoing

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2

ICU Medical, Inc.

Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

May 2, 2025 · DeviceView details →