Class IIOngoing
Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
December 12, 2025 · DeviceView details →