Micromed S.p.A.
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
Medline Industries, LP
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.