  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 841–860 of 1,000 recalls

Class IIOngoing

Allura Xper FD10/10; Model Number: 722005;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20; Model Number: 722012;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.

Diagnostica Stago, Inc.

Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20/10; Model Number: 722029;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

ICU Medical, Inc.

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD10; Model Number: 722010;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17

Sysmex America, Inc.

Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20 Biplane; Model Number: 722008;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Plum Duo Infusion System, List Number 40002-04-01

ICU Medical, Inc.

ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.

December 15, 2025 · DeviceView details →
Class IIOngoing

Brevera Breast Biopsy System Disposable 9 Gauge Needle. Biopsy Instrument.

Hologic, Inc

Presence of particulates in affected devices that can be deposited into breast tissue during use.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20 OR Table; Model Number: 722023;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20 Biplane; Model Number: 722013;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10

Wilson-Cook Medical Inc.

Nasal feeding tube packaged without the nasal transfer tube component.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD10; Model Number: 722026;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Allura Xper FD20; Model Number: 722006;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

December 15, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No

Elekta, Inc.

Due to a manufacturing issue and electrical grounding of systems.

December 14, 2025 · DeviceView details →
Class IIOngoing

Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

Integra LifeSciences Corp. (NeuroSciences)

Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

December 12, 2025 · DeviceView details →
Class IIOngoing

Medline medical procedure convenience kits labeled as: 1) WATKINS SPINE PACK-LF, Kit SKU DYNJ0275635M; 2) SM-SPINE PACK-LF, Kit SKU DYNJ0395037X; 3) TOTAL HIP PACK, Kit SKU DYNJ04809N; 4) MAIN/LAMINECTOMY PACK-LF, Kit SKU DYNJ0565499AN; 5) SCC/SPOC HAND & FOOT PACK, Kit SKU DYNJ27164Q; 6) ORTHO SPINE, Kit SKU DYNJ40221G; 7) URO GYN PACK, Kit SKU DYNJ45311J; 8) LAMINECTOMY PACK, Kit SKU DYNJ46330C; 9) DJ ORTHO SPINE, Kit SKU DYNJ48980M; 10) CERVICAL PACK, Kit SKU DYNJ52995F; 11) NORTH BACK FUSION PACK-LF, Kit SKU DYNJ56483L; 12) KNEE ARTHROSCOPY CHRISTUS, Kit SKU DYNJ61200B; 13) KNEE ARTHROSCOPY PACK-LF, Kit SKU DYNJ66150; 14) HIP PACK, Kit SKU DYNJ66630; 15) OHNS FREE FLAP A PACK, Kit SKU DYNJ69065F; 16) LUMBAR PACK, Kit SKU DYNJ69113C; 17) LAMINECTOMY PACK, Kit SKU DYNJ69401D; 18) LAMINECTOMY PACK, Kit SKU DYNJ81013D; 19) HIP/SHOULDER ARTHROSCOPY PACK, Kit SKU DYNJ84275A; 20) FLASC ACL PACK, Kit SKU DYNJ88971; 21) SPINE PACK MERCY GALENA, Kit SKU DYNJ89943; 22) SPINAL FUSION MT CARMEL, Kit SKU DYNJ900173J; 23) HAND, Kit SKU DYNJ904945G; 24) LAMINECTOMY, Kit SKU DYNJ909505D; 25) SHOULDER SPLIT, Kit SKU DYNJ910543F; 26) SHOULDER SPLIT, Kit SKU DYNJ910543G; 27) CRANIOTOMY PACK, Kit SKU DYNJT4724; 28) SPINE PACK, Kit SKU DYNJT5184.

Medline Industries, LP

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

December 12, 2025 · DeviceView details →
Class IIOngoing

Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.

Elekta, Inc.

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

December 12, 2025 · DeviceView details →
Class IIOngoing

Medline medical procedure convenience kits labeled as: NERVE BLOCK TRAY 2, Kit SKU DYNJRA2713

Medline Industries, LP

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

December 12, 2025 · DeviceView details →
← PreviousPage 43 of 50Next →