  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 661–680 of 1,000 recalls

Class IITerminated

Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630

Paragon 28, Inc.

A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.

July 30, 2021 · DeviceView details →
Class IITerminated

Achieva 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781277 (OUS)

Philips North America Llc

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

July 30, 2021 · DeviceView details →
Class IITerminated

Ingenia 1.5T Evoluton- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781315

Philips North America Llc

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

July 30, 2021 · DeviceView details →
Class IITerminated

Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781342

Philips North America Llc

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

July 30, 2021 · DeviceView details →
Class IITerminated

Arjo Disposable Repositioning Sling, Model AHD001

Getinge Dominican Republic SA

Sling loop straps on device pose a trip hazard.

July 30, 2021 · DeviceView details →
Class IITerminated

Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS)

Philips North America Llc

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

July 30, 2021 · DeviceView details →
Class IITerminated

Ingenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781357 (OUS)

Philips North America Llc

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

July 30, 2021 · DeviceView details →
Class IITerminated

OASIS MRI System

Hitachi Healthcare Americas Corporation

There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.

July 29, 2021 · DeviceView details →
Class IITerminated

LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420

Lymol Medical Corporation

Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.

July 29, 2021 · DeviceView details →
Class IITerminated

ECHELON MRI System

Hitachi Healthcare Americas Corporation

There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.

July 29, 2021 · DeviceView details →
Class IITerminated

ECHELON Oval MRI System

Hitachi Healthcare Americas Corporation

There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.

July 29, 2021 · DeviceView details →
Class IITerminated

Corin Unity Knee CS Insert, Size 5 / 9.0mm, Fixed Condylar Stabilised - UHMWPE, REF 1125182, Sterile, and Corin Unity Knee CS Insert, Size 8 / 10.0mm, Fixed Condylar Stabilised - UHMWPE, REF 1125243, Sterile, both are components of the Unity Total Knee System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin BIOLOX Delta Mod Head, diameter 28mm Medium 0mm, 12/14 Taper, REF 104.2805, Sterile, a component of the Trinity Acetabular System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin MetaFix Hip Stem, Size 1, Short Neck, 135 degrees, Collared Hip Stem, REF 579.0201, Sterile.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin TriFit TS, Size 8, Standard Tapered Stem, REF 694.0008, Sterile.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin Unity Knee Patella, Size 3 / 8.5mm, Centered Dome 3 PEG-UHMWPE, REF 112.018.06, Sterile, a component of the Unity Total Knee System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin Unity Knee Augments, Size 5, 6 & 7 / 10mm Thick, Femoral Posterior Augment, REF 1122370, Sterile, a component of the Unity Total Knee System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin Trinity-i, diameter 50mm Taper Size 3, Non-Occluded Ti Plasma Shell, REF 320.03.350, Sterile, a component of the Trinity Acetabular System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin Trinity Liner ECIMA, diameter 40mm Neutral Offset, Taper Size 4, REF 322.04.640, Sterile; and Corin Trinity Liner ECIMA, diameter 36mm Neutral 4mm EPW, Taper Size 4 (Elevated Wall), REF 322.04.936, Sterile; both are components of the Trinity Acetabular System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
Class IITerminated

Corin Trinity Dual Mobility, Size 4, Dual Mobility CoCr Liner, REF 321.04.540, Sterile, and Corin Dual Mobility, Size 4 42 insert/28 head, REF 325.04.042, Sterile; both are components of the Trinity Dual Mobility System.

Corin Ltd

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

July 28, 2021 · DeviceView details →
← PreviousPage 34 of 50Next →