Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.
Showing 541–560 of 1,000 recalls
Medline Industries, LP
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Tosoh Bioscience Inc
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.