Murata Vios, Inc.
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
Philips Healthcare
In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.