  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 261–280 of 1,000 recalls

Class IIOngoing

Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F

Medline Industries, LP

Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.

July 31, 2026 · DeviceView details →
Class IIOngoing

Percutaneous Insertion Kit (PIK), dilator set L, Model Numbers: 701054427 (USA), 701054427 (OUS);

Maquet Cardiopulmonary Gmbh

The packaging may not consistently maintain sterile barrier integrity.

July 31, 2026 · DeviceView details →
Class IIOngoing

Convenience kits CATARACT PACK DYNJ59881A

Medline Industries, LP

Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.

July 31, 2026 · DeviceView details →
Class IIOngoing

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Datex-Ohmeda, Inc.

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

July 30, 2026 · DeviceView details →
Class IIOngoing

Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Datex-Ohmeda, Inc.

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

July 30, 2026 · DeviceView details →
Class IIOngoing

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

GE Medical Systems Information Technologies Inc

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

July 30, 2026 · DeviceView details →
Class IIOngoing

Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000

Datex-Ohmeda, Inc.

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

July 30, 2026 · DeviceView details →
Class IIOngoing

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR INSERT 50mm OD x 28mm ID 10DEGREE HOOD C-M071-8-400 BEARING AND AXLE BUSHING ASSY STANDARD ELBOW C-M110-1-201 BUSHING AND BEARING USE C-MM100253-00 C-MM100253-02 DURACON TIB INSERT LG 12MM MED 14MM LAT C-M066-3-000 EXETER WINGLESS CENTRALISER FOR C-M082-7-302 C-M082-7-303 EXETER WINGLESS CENTRALIZER XSM C-M107-5-301 GMRS EXTRA SM ALL-POLY TIBIA SIZE 11mm C-MM100616-02 GMRS EXTRA SM ALL-POLY TIBIA SIZE 16mm C-MM100616-03 GMRS EXTRA SM ALL-POLY TIBIA SIZE 8mm C-MM100616-01 GMRS SML POLY TIBIA 11MM CASE 100520 C-MM100520-03 GMRS SML POLY TIBIA 16MM CASE 100520 C-MM100520-04 GMRS SML POLY TIBIA 8MM CASE 100520 C-MM100520-02 GMRS SML TIBIA 11MM WITH CASE 100411 C-MM100411-03 GMRS SML TIBIA 14MM WITH CASE 100411 C-MM100411-04 GMRS SML TIBIA 8MM WITH CASE 100411 C-MM100411-02 GMRS TIB INSRT SML 10MM WITH CASE 100507 C-MM100507-04 GMRS TIB INSRT SML 16MM WITH CASE 100507 C-MM100507-06 KALI TIBIAL INSERT LARGE RIGHT 8mm C-M071-6-601 KRH REV TIB INS 16mm THK .655 DIA STEM C-M113-5-603 MRH KNEE TIBIAL INSERT FOR THIN POST C-M067-3-103 MRS RADIAL BUSHING FORITH C-M107-7-800 C-M107-7-801 SELF-PRESSURIZING GLENOID SIZE 9 C-M064-4-300 TRIATHLON NO. 9 TIBIAL INSERT 11MM C-M067-5-003 TRIATHLON NO. 9 TIBIAL INSERT 9MM C-M067-5-002 TRIDENT ALL POLY CUP 44 X 66MM C-M066-0-901 TRIDENT CONSTRAINED ACETABULAR INSERT SIZE A X 22mm ID UH1 C-M104-2-002 TRIDENT CONSTRAINED ACETABULAR INSERT SIZE B X 22mm ID UH1 C-M104-2-003 UHR BIPOLAR 22X38MM UH1-38-22 UHR BIPOLAR 22X40MM UH1-40-22 UHR BIPOLAR 22X41MM UH1-41-22 UHR BIPOLAR 22X42MM UH1-42-22 UHR BIPOLAR 22X43MM UH1-43-22 UHR BIPOLAR 22X45MM UH1-45-22 UHR BIPOLAR 26X41mm UH1-41-26 UHR BIPOLAR 26X42mm UH1-42-26 UHR BIPOLAR 26X43mm UH1-43-26 UHR BIPOLAR 26X44mm UH1-44-26 UHR BIPOLAR 26X45mm UH1-45-26 UHR BIPOLAR 26X46mm UH1-46-26 UHR BIPOLAR 26X47mm UH1-47-26 UHR BIPOLAR 26X48mm UH1-48-26 UHR BIPOLAR 28X42mm UH1-42-28 UHR BIPOLAR 28X44mm UH1-44-28 UHR BIPOLAR 28X45mm UH1-45-28 UHR BIPOLAR 28X46mm UH1-46-28 UHR BIPOLAR 28X47mm UH1-47-28 UHR BIPOLAR 28X48mm UH1-48-28 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT 48MM 69-2248 ALL POLY CONSTRAINED INSERT 50MM 69-2850 ALL POLY CONSTRAINED INSERT 52MM 69-2852 CONSTRAINED ACETABULAR INSERTS 22MM 50MM 2099-2250 CONSTRAINED ACETABULAR INSERTS 22MM 52MM 2099-2252 CONSTRAINED ACETABULAR INSERTS 22MM 56MM 2099-2256 CONSTRAINED ACETABULAR INSERTS 28MM 61MM 2099-2861 EIUS UNI TIB SMALL RM/LL 9MM 6636-0-419 PCA CONST LINER 32MM NEUTRAL C - SHELL 6299-N-332 SYS 12 CONST LINER 32ID 10 DEG P6 +4MM 6302-H-632 TRIDENT 0 DEG CONSTRAINED INSERT 22D 690-00-22D TRIDENT 0 DEG CONSTRAINED INSERT 22E 690-00-22E TRIDENT 0 DEG CONSTRAINED INSERT 28F 690-00-28F TRIDENT 0 DEG CONSTRAINED INSERT 28G 690-00-28G TRIDENT 0 DEG CONSTRAINED INSERT 32I 690-00-32I TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D TRIDENT CONSTR ACET INSERT 690-10-28K TRIDENT CONSTRAINED INSERT 690-10-22E 690-10-22F 690-10-28G 690-10-28H 690-10-28I Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PATELLA 73-2710 SCORPIO INSET PATELLA 73-2610 SCORPIO M-DOME PATELLA 73-0110 73-0710 73-0910 SCORPIO RECESSED PATELLA 26MM 3044-0026 SCORPIO TS TIB INS 72-4-9114 SCORPIO TS TIBIAL INSERT 72-4-0324 SCORPIOFLEX TOTAL KNEE CR 72-16-0512 72-16-0518 72-16-0912 72-16-1112 SERIES II TIBIAL BEAR INSERT NO 11 15MM 3052-1115 UHR BIPOLAR 26X52mm UH1-52-26 UHR BIPOLAR 26X58mm UH1-58-26 UHR BIPOLAR 28X48mm UH1-48-28 UHR BIPOLAR 28X52mm UH1-52-28 UHR BIPOLAR 28X55mm UH1-55-28 UHR BIPOLAR 32X49MM UH1-49-32 UHR BIPOLAR 32X51MM UH1-51-32 UHR BIPOLAR 32X52MM UH1-52-32 UHR BIPOLAR 32X54MM UH1-54-32 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER - GLENOID #11 5361-6111 SHOULDER - GLENOID #5 5361-6105 SHOULDER - GLENOID #7 5361-6107 SHOULDER - GLENOID #9 5361-6109 SHOULDER - KEELED GLENOID #11 5361-6211 SHOULDER - KEELED GLENOID #5 5361-6205 SHOULDER - KEELED GLENOID #7 5361-6207 SHOULDER - KEELED GLENOID #9 5361-6209 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
Class IIOngoing

Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000

Datex-Ohmeda, Inc.

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

July 30, 2026 · DeviceView details →
Class IIOngoing

AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Howmedica Osteonics Corp.

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

July 30, 2026 · DeviceView details →
← PreviousPage 14 of 50Next →