  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 81–100 of 1,000 recalls

Class IOngoing

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Baxter Healthcare Corporation

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

May 21, 2026 · DeviceView details →
Class IOngoing

Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device

Baxter Healthcare Corporation

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

May 21, 2026 · DeviceView details →
Class IOngoing

Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).

Insulet Corporation

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

May 20, 2026 · DeviceView details →
Class IOngoing

Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).

Insulet Corporation

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

May 20, 2026 · DeviceView details →
Class IOngoing

Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).

Insulet Corporation

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

May 20, 2026 · DeviceView details →
Class IOngoing

Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

Abiomed, Inc.

Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

May 18, 2026 · DeviceView details →
Class IOngoing

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.

Medline Industries, LP

Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

May 15, 2026 · DeviceView details →
Class IOngoing

Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 5. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 6. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 7. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 8. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 9. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 10. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 11. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 12.Impella Optical Controller, Packaged, US, UK; Product Code: 0042-0010-US. 13. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 14. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 15. Optical AIC w/Impella Connect, Pack'd, JP, EU; Product Code: 0042-0040-JP. 16. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU. 17. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 18. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.

Abiomed, Inc.

Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

May 14, 2026 · DeviceView details →
Class IOngoing

Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds

Medical Action Industries, Inc. 306

Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels.

May 13, 2026 · DeviceView details →
Class IOngoing

Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance

Datex-Ohmeda Inc

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

May 8, 2026 · DeviceView details →
Class IOngoing

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Datex-Ohmeda Inc

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

May 8, 2026 · DeviceView details →
Class IOngoing

Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Datex-Ohmeda Inc

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

May 8, 2026 · DeviceView details →
Class IOngoing

Atlan A350XL. Model Number: 8621600. anesthesia workstation

Draeger, Inc.

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

May 7, 2026 · DeviceView details →
Class IOngoing

Atlan A350. Model Number: 8211500. anesthesia workstation

Draeger, Inc.

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

May 7, 2026 · DeviceView details →
Class IOngoing

Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333201); Spinocan 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333211); Spinocan 22 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333231)

B Braun Medical Inc

These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.

May 6, 2026 · DeviceView details →
Class IOngoing

Medline Convenience Kits: 1) DYNJ905503F, Model Number: CV ANESTHESIA - ROOM SET UP; 2) DYNJ905503G, Model Number: CV ANESTHESIA - ROOM SET UP

Medline Industries, LP

The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.

May 6, 2026 · DeviceView details →
Class IOngoing

Medline Convenience Kits: 1) DRAWER 6A ADULT CENTRAL LINE, Model Number: ACC010890; 2) OPEN HEART A OVERHEAD, Model Number: DYNJ910066; 3) OPEN HEART, Model Number: DYNJ9855016V; 4) OPEN HEART, Model Number: DYNJ9855016X

Medline Industries, LP

The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.

May 6, 2026 · DeviceView details →
Class IOngoing

Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); 25GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333862); GOVP24BK 24G PENCAN TRAY W/ BUP & DEXTROS (332607); P24BK 24G PENCAN TRAY W/BUP & DEXTROSE (333868); P25BKAY PENCAN SPINAL TRAY (333865); P27BK PENCAN SPINAL W/BUPIVAC (333871); P25BK SPINAL TRAY (333851)

B Braun Medical Inc

These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.

May 6, 2026 · DeviceView details →
Class IOngoing

Medline Convenience Kits: 1) DRAWER 6 CENTRAL LINE INSERTIO, Model Number: ACC011142; 2) CENTRAL AND ARTERIAL LINE TOTE, Model Number: DYNJ8407

Medline Industries, LP

The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.

May 6, 2026 · DeviceView details →
Class IOngoing

Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840144Q; 2) OPEN HEART, Model Number: CDS840150W; 3) OPEN HEART ADULT, Model Number: CDS840396AA; 4) CABG A CDS, Model Number: CDS982269N; 5) OPEN HEART ANESTHESIA PACK, Model Number: CDS983849G; 6) OPEN HEART ACCESSORY CDS, Model Number: CDS984289L; 7) OPEN HEART A, Model Number: CDS985105Q; 8) OPEN HEART PART 2, Model Number: DYNJ901075R; 9) VPH OPEN HEART, Model Number: DYNJ902664L; 10) HEART A, Model Number: DYNJ903465I; 11) HEART, Model Number: DYNJ903822F; 12) OPEN HEART BASIC B, Model Number: DYNJ904311F; 13) SM OPEN HEART PACK A&B, Model Number: DYNJ904749J; 14) OPEN HEART ANESTHESIA ICC-B, Model Number: DYNJ905567B; 15) SILVER CROSS OPEN HEART, Model Number: DYNJ906102K; 16) ANESTHESIA OPEN HEART, Model Number: DYNJ906225I; 17) OPEN HEART COMBO, Model Number: DYNJ907400F; 18) CARDIAC CABG B, Model Number: DYNJ908121C; 19) FS OPEN HEART, Model Number: DYNJ908308A; 20) OPEN HEART SUPPLEMENT, Model Number: DYNJ909116D; 21) MEM CARDIAC, Model Number: DYNJ9622573AP; 22) OPEN HEART A & B PACK, Model Number: DYNJ9718008AT

Medline Industries, LP

The Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.

May 6, 2026 · DeviceView details →
← PreviousPage 5 of 50Next →