Civic Data
FDA RecallsUSDA RecallsAbout

FDA Recalls

  • Food recalls
  • Drug recalls
  • Device recalls

USDA Recalls

  • All USDA recalls
  • Poultry recalls
  • Meat recalls
  • Egg recalls

About

  • About us
  • FDA guides
  • USDA guides
  • Data sources
  • Privacy policy

Data

  • OpenFDA API
  • USDA FSIS

© 2026Civic Data · Data sourced from OpenFDA and USDA FSIS · Not affiliated with the U.S. Food and Drug Administration or the U.S. Department of Agriculture · Always verify against official sources before acting

  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 901–920 of 1,000 recalls

Class ITerminated

NeuroBlate System SideFire Select 2.2 mm Directional Laser Probe, Sterile EO. Model #s SFS000-01,SFS002-01, SFS122-01, SFS222-01, SFS322-01, SFS422-01, and SFS522-01. NeuroBlate System SideFire Select Directional Laser Probes are part of the NeuroBlate System. The Monteris Medical NeuroBlate System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It also provides real-time thermographic analysis of selected MRI images. The laser delivery probes direct the laser energy to the proper treatment location.

Monteris Medical Corp

Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.

March 25, 2016 · DeviceView details →
Class ITerminated

Boston Scientific Fetch 2 Aspiration Catheter, Catalog No. 109400-001. Product The Fetch 2 Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature.

Boston Scientific Corporation

Boston Scientific is recalling all models of Fetch 2 Aspiration Catheter as a result of receiving reports from the field of catheter shaft breaks; including cases requiring the use of a snare or other intervention to retrieve the http://webapps.cdrh.fda.gov/division-tracking/images/trash.pngbroken shaft from the patient.

March 22, 2016 · DeviceView details →
Class ITerminated

Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-pressure seal. Closure of the high-pressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and allows for pressure injections up to 600 psi.

Vascular Solutions, Inc.

Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.

March 4, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P640 REF 418579), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P645 REF 419602), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P639 REF 418662), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P644 REF 418673), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P642 REF 418604), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P646 REF 420144), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P636 REF 417951), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P641 REF 418590), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P643 REF 418671), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Biomerieux Inc

Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.

February 24, 2016 · DeviceView details →
Class ITerminated

Dexcom G4 PLATINUM (Professional) Receiver Receiver Part Number: MT20649 Receiver (mg/dL), Black. Intended to detect trends and track glucose patterns in persons with diabetes. The G4 PLATINUM CGM System and G4 PLATINUM (Professional) CGM System use the same receiver, MT20649. The kits differ due to other kit contents, such as indications for use and professional receiver shield.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Dexcom G4 PLATINUM (Pediatric) Receiver Receiver Part Number: MT22430 Receiver (mg/dL), Black (pediatrics); MT22430-PNK, Receiver (mg/dL), Pink (pediatrics); MT22430-BLU, Receiver (mg/dL); Blue (pediatrics). Intended to detect trends and track glucose patterns in persons with diabetes.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Dexcom G4 PLATINUM Receiver Receiver Part Number: MT20649 Receiver (mg/dL), Black; MT20649-PNK, Receiver (mg/dL), Pink; MT20649-BLU, Receiver (mg/dL), Blue. Intended to detect trends and track glucose patterns in persons with diabetes.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Dexcom G5 Mobile Receiver Receiver Part Number: MT22719 Receiver (mg/dL), Black; MT22719-PNK, Receiver (mg/dL), Pink; MT22719-BLU, Receiver (mg/dL), Blue. Intended to detect trends and track glucose patterns in persons with diabetes.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Dexcom G4 PLATINUM (Pediatric) Receiver with Share Receiver Part Number: MT22608 Receiver (mg/dL), Black (pediatrics); MT22608-PNK, Receiver (mg/dL), Pink (pediatrics); MT22608-BLU, Receiver (mg/dL), Blue (pediatrics). Intended to detect trends and track glucose patterns in persons with diabetes.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Dexcom G4 PLATINUM Receiver with Share Receiver Part Number: MT22495 Receiver (mg/dL), Black; MT22495-PNK, Receiver (mg/dL), Pink; MT22495-BLU, Receiver (mg/dL), Blue. Intended to detect trends and track glucose patterns in persons with diabetes.

Dexcom Inc

Customers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.

February 23, 2016 · DeviceView details →
Class ITerminated

Percutaneous Insertion Tray; Product Code: IAK-06845 The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Arrow International, Inc., Division of Teleflex Medical Inc.

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

February 10, 2016 · DeviceView details →
Class ITerminated

Ultra 8 IAB: 8Fr 40cc; Product Code: IAB-05840-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.

Arrow International, Inc., Division of Teleflex Medical Inc.

The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.

February 10, 2016 · DeviceView details →