OriGen Biomedical, Inc.
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
OriGen Biomedical, Inc.
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.