  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 801–820 of 1,000 recalls

Class IOngoing

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: 6.0MM FLEXTEND PEDIATRIC PLUS VEE TRACH TUBE , Product Code/List Number/Item Code 60PFP60

Smiths Medical ASD Inc.

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

May 29, 2024 · DeviceView details →
Class IOngoing

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM AIRE-CUF , Product Code/List Number/Item Code 65P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM AIRE-CUF , Product Code/List Number/Item Code 65P030; c) PAED, TRACHEOSTOMY TUBE 3.5mm AIRE-CUF , Product Code/List Number/Item Code 65P035; d) PAED. TRACHEOSTOMY TUBE 4.0MM AIRE-CUF , Product Code/List Number/Item Code 65P040; e) PAED. TRACHEOSTOMY TUBE 4.5MM AIRE-CUF , Product Code/List Number/Item Code 65P045; f) PAED. TRACHEOSTOMY TUBE 5.0MM AIRE-CUF , Product Code/List Number/Item Code 65P050; g) PAED. TRACHEOSTOMY TUBE 5.5MM AIRE-CUF , Product Code/List Number/Item Code 65P055;

Smiths Medical ASD Inc.

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

May 29, 2024 · DeviceView details →
Class IOngoing

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P025; b) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P030; c) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P035; d) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P040; e) TTS PEDIATRIC TRACH TUBE, Product Code/List Number/Item Code 67P045; f) TTS PEDIATRIC TRACH TUBE, Product Code/List Number/Item Code 67P050; g) TTS PEDIATRIC TRACH TUBE, Product Code/List Number/Item Code 67P055

Smiths Medical ASD Inc.

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

May 29, 2024 · DeviceView details →
Class IOngoing

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP40; b) PEDIATRIC TRACHEOSTOMY TUBE 4.5mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP45; c) PEDIATRIC TRACHEOSTOMY TUBE 5.0mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP50; d) PEDIATRIC TRACHEOSTOMY TUBE 5.5mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP55; e) PEDIATRIC TRACHEOSTOMY TUBE 6.0mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP60

Smiths Medical ASD Inc.

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

May 29, 2024 · DeviceView details →
Class IOngoing

PORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM UNCUFFED NEONATAL FLEXTEND PLUS TRACHEOSTOMY TUBE W/NEO STRAIGHT STRAP, Product Code/List Number/Item Code 60NFPS25

Smiths Medical ASD Inc.

Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.

May 29, 2024 · DeviceView details →
Class IOngoing

Inspire Model 3028, IV Implantable Pulse Generator

Inspire Medical Systems Inc.

There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,

May 16, 2024 · DeviceView details →
Class IOngoing

Ventilator HAMILTON-C6, PN: 160021

Hamilton Medical AG

Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.

May 15, 2024 · DeviceView details →
Class IOngoing

MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. Reusable Patient Return Electrode for patients weighing (0.8lb-50lb), (350g-22.7Kg), 66cm Long x 30.5cm Wide x 1.25cm. Electrosurgical use.

Megadyne Medical Products, Inc.

Reports of patient burns.

May 8, 2024 · DeviceView details →
Class IOngoing

Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device

Thoratec Corp.

System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected

May 8, 2024 · DeviceView details →
Class IOngoing

Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 11973. OUS units marketed with the following descriptions: Plum A+ Mednet TM Firmware New 13.6 Span, Plum A+3, PLUM A+3 INFUSION PUMPS and Plum A+3 Wireless Refurb, PLUM A+, PLUM A+ Infusion System, PlumA+3IV Pump Multi 1' Set

ICU Medical Inc

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

May 7, 2024 · DeviceView details →
Class IOngoing

Plum 360/A+ spare batteries. Item Number: SUB0000864.

ICU Medical Inc

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

May 7, 2024 · DeviceView details →
Class IOngoing

Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following descriptions: Bomba de infusi¿n Plum 360 compatible con el software ICU Medical MedNet, Plum 360 Infusion System, PLUM 360, PLUM 360" Infuser, Plum 360 Infuser compatible with ICU Medical MedNet Software, Plum 360 Infuser compatible with ICU Medical MedNet Software Refurb, Plum 360 sistema infusionale compatibile con ICU Medical MedNet Software, Pompe a perfusion Plum 360, compatible avec le logiciel ICU Medical MedNet

ICU Medical Inc

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

May 7, 2024 · DeviceView details →
Class IOngoing

Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.

Getinge Usa Sales Inc

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

May 1, 2024 · DeviceView details →
Class IOngoing

ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2-02-04 4. 799-EGL2-02-05 5. 799-EGL2-02-08 6. 799-EGL2-02-10 7. 799-EGL2-02-11 8. 799-EGL2-02-15 9. 799-EGL2-02-16 10. 799-EMBP-05-01 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 20. 799-EMVP-05-01 21. 799-EMVP-05-05 22. 799-EMVP-05-10 23. 799-EMVP-05-15 24. 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16

ZOLL Medical Corporation

Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy

April 30, 2024 · DeviceView details →
Class IOngoing

Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS

ARROW INTERNATIONAL Inc.

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

April 29, 2024 · DeviceView details →
Class IOngoing

Arrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U

ARROW INTERNATIONAL Inc.

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

April 29, 2024 · DeviceView details →
Class IOngoing

Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U

ARROW INTERNATIONAL Inc.

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

April 29, 2024 · DeviceView details →
Class IOngoing

Optum Nimbus II Plus, Model number AAA-00170; Ambulatory Infusion Pump

OptumHealth Care Solutions LLC

Nimbus II Plus, Infusion Pump Systems, manufactured by InfuTronix, are being removed the from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing.

April 18, 2024 · DeviceView details →
Class IOngoing

Medline procedure kits labeled as: 1) NEPHROLOGY PACK, Pack Number DYNJ59089; 2) NERVE BLOCK TRAY, Pack Number DYNJRA1221A

MEDLINE INDUSTRIES, LP - Northfield

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

April 8, 2024 · DeviceView details →
Class IOngoing

Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX. The TwinTube sample line 240cm is connected to the adapter for DVT. Twin Tube Probenschlauch (707004), which is sold outside the US only. Used in breath analysis for metabolic response study.

Vyaire Medical, Inc.

Potential of the nozzle separating during patient use.

April 8, 2024 · DeviceView details →
← PreviousPage 41 of 50Next →