Civic Data
FDA RecallsUSDA RecallsAbout

FDA Recalls

  • Food recalls
  • Drug recalls
  • Device recalls

USDA Recalls

  • All USDA recalls
  • Poultry recalls
  • Meat recalls
  • Egg recalls

About

  • About us
  • FDA guides
  • USDA guides
  • Data sources
  • Privacy policy

Data

  • OpenFDA API
  • USDA FSIS

© 2026Civic Data · Data sourced from OpenFDA and USDA FSIS · Not affiliated with the U.S. Food and Drug Administration or the U.S. Department of Agriculture · Always verify against official sources before acting

  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 781–800 of 1,000 recalls

Class IOngoing

MIC* Gastric-Jejunal Feeding Tube, Endoscopic/Radiologic Placement, GJ-Tube: a) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16; b) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-15; c) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-22; d) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-30; e) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18; f) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18-22; g) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18-30; h) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 22 Fr, Product Code 0250-22; i) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 16 Fr, Product Code 0260-16; j) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 18 Fr, Product Code 0260-18; k) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 22 Fr, Product Code 0260-22;

Avanos Medical, Inc.

MIC* Gastric - Jejunal Feeding Tube Kits contain a sterile pre-filled syringe subsequently recalled by Nurse Assist.

January 16, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004

Vyaire Medical

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

January 10, 2024 · DeviceView details →
Class IOngoing

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Ventec Life Systems, Inc.

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

January 9, 2024 · DeviceView details →
Class IOngoing

1. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3. HeartMate Touch Communication System, HMT1150 The HeartMate Touch Application is loaded onto a tablet and will provide a user interface with state-of-the-art hardware/software platforms for supporting LVAD patients. Along with being able to program and monitor the LVAS, the HeartMate Touch Application introduces features that enables users to setup pairing and connection with the existing HeartMate 3 Left Ventricular Assist System (LVAS) via the HeartMate Touch Wireless Adapter and generate reports from the log files that are downloaded from the HM3 System Controller.

Thoratec Corp.

Due to software and controller systems that results in unexpected pump start or pump stop.

January 3, 2024 · DeviceView details →
Class IOngoing

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002

Datex-Ohmeda, Inc.

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

December 29, 2023 · DeviceView details →
Class IOngoing

EvAir CPRSR Kit Upgrade: a) DISS/CEE 7/7 230V 50/60 HZ, Model Number M1230849; b) DISS/NEMA 5-15 115V 60 HZ, Model Number M1230847; c) NIST/BS 1363 230V 50/60 HZ, Model Number M1230849; d) NIST/BS 546 230V 50/60 HZ, Model Number M1230849; e) NIST/CEE 7/7 230V 50/60 HZ, Model Number M1230849;

Datex-Ohmeda, Inc.

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

December 29, 2023 · DeviceView details →
Class IOngoing

EVair air compressor, model numbers: M1230849 and M1230847

Datex-Ohmeda, Inc.

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

December 29, 2023 · DeviceView details →
Class IOngoing

Cardinal Health Monoject, 1 mL Purple Enteral Syringe with Enfit Connection Sterile, REF 401SE

Cardinal Health 200, LLC

Cardinal Health is expanding their previous product correction actions to a product removal of all sizes of Cardinal Health brand Monoject sterile Syringe Luer-Lock (1, 3, 6, 12, 20, 35, and 60 mL) and Cardinal Health brand Monoject sterile Enteral Syringes with the ENFit connection (1, 3, 6, 12, 35, and 60 mL) due to a change in manufacturing and rebranding efforts.

December 28, 2023 · DeviceView details →
Class IOngoing

Cardinal Health Monoject, 35mL Enteral Syringe (purple) - For Use with ENFit Connection System, REF 435SE

Cardinal Health 200, LLC

The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.

December 28, 2023 · DeviceView details →
Class IOngoing

Cardinal Health Monoject, 20 mL Syringe Luer-Lock Tip Soft Pack, REF 1182000777

Cardinal Health 200, LLC

Cardinal Health is expanding their previous product correction actions to a product removal of all sizes of Cardinal Health brand Monoject sterile Syringe Luer-Lock (1, 3, 6, 12, 20, 35, and 60 mL) and Cardinal Health brand Monoject sterile Enteral Syringes with the ENFit connection (1, 3, 6, 12, 35, and 60 mL) due to a change in manufacturing and rebranding efforts.

December 28, 2023 · DeviceView details →
Class IOngoing

Cardinal Health Monoject, 12 mL Syringe Luer-Lock Tip Soft Pack, REF 1181200777T

Cardinal Health 200, LLC

Cardinal Health is expanding their previous product correction actions to a product removal of all sizes of Cardinal Health brand Monoject sterile Syringe Luer-Lock (1, 3, 6, 12, 20, 35, and 60 mL) and Cardinal Health brand Monoject sterile Enteral Syringes with the ENFit connection (1, 3, 6, 12, 35, and 60 mL) due to a change in manufacturing and rebranding efforts.

December 28, 2023 · DeviceView details →
Class IOngoing

Cardinal Health Monoject, 6ML Enteral Syringe (purple) Low Dose (LDT) - For Use with ENFit Connection System, REF 406SE

Cardinal Health 200, LLC

The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.

December 28, 2023 · DeviceView details →