  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 661–680 of 1,000 recalls

Class ITerminated

Greatbatch, Offset Cup Impactor with the following Parts and Catalogue #: Tl0600 ~ MMP3100CHA04 T10491 MIMP3100CHA04 T12230 M2M1P3100CHA04 T15752 ~MMP3100CHA04 T15949 ~MMP3100CHA04 T17238 MMP3100CHA04 T9954 MMP3100CHA04 T9955 MMP3100CHA04 T8468 MMP3100CHA04 T11506 MMP3100CHA13 T10281 MMP3100CHA04 T8044 ~MMP3100CHA04 T17203 87-4 20-00980 T8177 ~ MMP3100CHA04 T16934 MMP3100CHA62 Tll340 SHP9109.00010 T8043 ~ MMP3100CHA04 T17321 MMPP3100CHA04 T15311 GG920010029 T1 7650 255000110 T8143 803-t 1-103 Tl4384 EK-00200690 T1666t XEB227242 T9316 MIMP3100CHA04 Tl3642 500325 T10753 ~MMP3100CH04 T9348 ~ MMP3100CHA04 T13480 MMP3100CHA042 T10243 ~MIMP3001CHA08 Tll209 ~ MMP300CHA12 Tl0861 MMP3100CHA04 T16427 MMP3100CHA04 T9894 :-. MMP3100CHA04 T12861 MMP3100CHA\041 T8088 2002CUPI T1661 1 T16611 Standard Offset Cup Impactors are handheld, reusable surgical instruments used to position and seat acetabular cups into the reamed acetabulum during hip joint replacement surgery. They are equipped with a suitable connection system to the acetabular cup (e.g. a thread) and can be supplied with a fixed or removable shock absorber. A hammer is used to deliver the necessary impact to an anvil (or impactor disc) at the rear end of the instrument.

Greatbatch Medical

On December 17, 2015, Greatbatch discovered that the Standard Offset Cup Impactor with a POM-C handle grip failed sterility testing when housed in a dedicated instrument case using the sterilization parameters identified in the Standard Offset Cup Impactors Instructions for Use (IFU).

August 2, 2016 · DeviceView details →
Class ITerminated

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

OraSure Technologies, Inc.

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

July 20, 2016 · DeviceView details →
Class ITerminated

Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio2 Replacement Monitor (Home) Model Number: 0200457 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: The Alere INRatio 2 PT/INR Home Monitoring System, consisting of the INRatio 2 Home Monitor and INRatio PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin therapy on clotting time by properly selected suitably trained users (by prescription for home use or other order of a treating physician). Patients must be stabilized (>6 weeks) on warfarin therapy. The Alere INRatio¿2 PT/INR Home Monitoring System is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿2 PT/INR Home Monitoring System is not intended to be used for screening purposes.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio2 PT/INR Professional Testing System Model Number: 0200431 Product Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio Replacement Monitor Model Number: 0100137 Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

INRatio Prothrombin Time (PT) Monitoring System Model Number: 0100007 Product Usage: The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test strip area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the Prothrombin time test on the blood sample. The other two channels contain the reagents. to perfom the Prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control test are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio2 PT/INR Home Monitoring System Model Number: 0200432, 0200433 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: The Alere INRatio 2 PT/INR Home Monitoring System, consisting of the INRatio 2 Home Monitor and INRatio PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin therapy on clotting time by properly selected suitably trained users (by prescription for home use or other order of a treating physician). Patients must be stabilized (>6 weeks) on warfarin therapy. The Alere INRatio¿2 PT/INR Home Monitoring System is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿2 PT/INR Home Monitoring System is not intended to be used for screening purposes.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use): The Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use), consisting of the INRatio ¿ 2 Monitor and INRatio¿ 2 PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals. The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is not intended to be used for screening purposes.

Alere San Diego, Inc.

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

July 11, 2016 · DeviceView details →
Class ITerminated

NucliSENS easyMAG Magnetic Silica The NucliSENS easyMAG accessory products are reagents and disposable to be used with the bioMerieux extraction systems which mention these products in their User Manuals. They enable the automated extraction (purification and concentration) of total nucleic acids (RNA/DNA) from biological samples which are liquid and homogeneous. These products are intended for in vitro diagnostic use by healthcare professionals only.

BioMerieux SA

Several customer complaints about amplification performance issues when using different batches of MagSil with the NucliSENS easyMAG extraction platform.

July 8, 2016 · DeviceView details →
Class ITerminated

DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or elevate organs and tissue to provide better visualization access.

Carefusion 2200 Inc

BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the potential to damages tissue or organs

July 1, 2016 · DeviceView details →
Class ITerminated

Light Adaptor for Small Battery Drive and Small Battery Drive II.

The Anspach Effort, Inc.

There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.

June 24, 2016 · DeviceView details →
Class ITerminated

Adaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in traumatology and orthopedic surgery of the skeleton, i.e. drilling, reaming or cutting bone. The Adaptors (05.001.024 and 05.001.108) can be used with one of the electrical consoles belonging to the Electric Pen Drive (EPD) to operate the Small Battery Drive and Small Battery Drive II.

The Anspach Effort, Inc.

There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.

June 24, 2016 · DeviceView details →
Class ITerminated

Roadrunner Uniglide Hydrophilic Wire Guide, Catalog Numbers and GPN: HPWA-18-260, G30489; HPWA-18-320, G30490; HPWA-25-260, G30491; HPWA-35-260, G30492; HPWAS-35-260, G30506; HPW-35-150, G56149; HPW-35-180, G56150; HPWS-35-150, G56152; HPWS-35-180, G56153; HPWA-35-80, G56172; HPWA-35-150, G56173; HPWA-35-180, G56174; HPWA-18-180, G56160; HPWAS-35-150, G56176; HPWAS-35-180, G56177. Used with vascular catheter. Roadrunner UniGlide Hydrophilic Wire Guide is constructed from a steerable, metallic core with a polymer coating. A hydrophilic coating is applied over the radiopaque polymer jacket. Sterile- Individual product is packaged in a Tyvek-film sterilizable outer pouch and boxed in a quantity of five.

Cook Inc.

Potential coating contamination with glass particles.

June 24, 2016 · DeviceView details →
Class ITerminated

ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA41, SIZE 4+; SURFACE POROUS, HA COATED; FIT CEMENTLESS; USE W/ SIZE 4 FEMUR & SIZE 4 TIBIAL INSERT; CONTENTS, 1 EACH, STERILE R, Rx ONLY, MicroPort Knee Systems Total Knee Systems MicroPort Total Knee Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

MicroPort Orthopedics Inc.

ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.

June 21, 2016 · DeviceView details →
Class ITerminated

ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA10, SIZE 1 STD; SURFACE POROUS, HA COATED; FIT CEMENTLESS; USE W/ SIZE 1 FEMUR & SIZE 1 TIBIAL INSERT; CONTENTS, 1 EACH, STERILE R, Rx ONLY, MicroPort Knee Systems Total Knee Systems MicroPort Total Knee Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

MicroPort Orthopedics Inc.

ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.

June 21, 2016 · DeviceView details →
Class ITerminated

ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA40, SIZE 4 STD; SURFACE POROUS, HA COATED; FIT CEMENTLESS; USE W/ SIZE 4 FEMUR & SIZE 4 TIBIAL INSERT; CONTENTS, 1 EACH, STERILE R, Rx ONLY, MicroPort Knee Systems Total Knee Systems MicroPort Total Knee Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

MicroPort Orthopedics Inc.

ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.

June 21, 2016 · DeviceView details →
Class ITerminated

ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA11, SIZE 1+; SURFACE POROUS, HA COATED; FIT CEMENTLESS; USE W/ SIZE 1 FEMUR & SIZE 1 TIBIAL INSERT; CONTENTS, 1 EACH, STERILE R, Rx ONLY, MicroPort Knee Systems Total Knee Systems MicroPort Total Knee Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

MicroPort Orthopedics Inc.

ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.

June 21, 2016 · DeviceView details →
← PreviousPage 34 of 50Next →