  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 601–620 of 1,000 recalls

Class IOngoing

McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx Only

Covidien

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

July 9, 2024 · DeviceView details →
Class IOngoing

McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx Only

Covidien

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

July 9, 2024 · DeviceView details →
Class IOngoing

Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100

Covidien

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

June 24, 2024 · DeviceView details →
Class IOngoing

NIM Vital Nerve Monitoring systems (all serial numbers manufactured) with NIM Vital System versions: v1.0.0, v1.1.1, v1.3.2, v1.4.3. NIM Vital Nerve Integrity Monitor with below product descriptions. 1. CONSOLE NIM4CM01 NIM 4.0, Model Number: NIM4CM01. 2. CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, Model Number: NIM4CM01RF, 3. PATIENT INTERFACE NIM4CPB1 NIM 4.0, Model Number: NIM4CPB1, 4. PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB, Model Number: NIM4CPB1RF, 5. SOFTWARE NIM4SWU143 UPGRADE V1.4.3, Model Number: NIM4SWU143

Medtronic Xomed, Inc.

Potential for false negative response on the NIM Vital Nerve Monitoring System

June 24, 2024 · DeviceView details →
Class IOngoing

Brand Name: MEGADYNE" MEGA SOFT" Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" Reusable Patient Return Electrode Model/Catalog Number: 0830 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega Soft, Single Cord, Adult, For Patients Weighing over 25lbs (11.4Kg), 117cm Long x 51cm Wide x 1.25cm Thick Component: No

Megadyne Medical Products, Inc.

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

June 17, 2024 · DeviceView details →
Class IOngoing

Brand Name: MEGADYNE" MEGA 2000" Patient Return Electrode Product Name: MEGADYNE" MEGA 2000" Patient Return Electrode Model/Catalog Number: 0800 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega 2000, Adult, For Patients Weighing over 25lbs. (11.4Kg), 91cm Long x 51cm Wide Component: No

Megadyne Medical Products, Inc.

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

June 17, 2024 · DeviceView details →
Class IOngoing

Brand Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Model/Catalog Number: 0835 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega Soft, Dual Cord, Adult, For Patients Weighing over 25lbs. (11.4kg), 117cm Long x 51cm Wide x 1.25cm Thick Component: No

Megadyne Medical Products, Inc.

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

June 17, 2024 · DeviceView details →
Class IOngoing

BAUSCH+LOMB, I-PACK Injection Kit, REF 18060, for the administration of intravitreal injection (eye)

Synergetics Inc

Sterilization certificates could not be validated by the supplier

June 12, 2024 · DeviceView details →
Class IOngoing

BAUSCH+LOMB, I-PACK Injection Kit with Drape, REF 18061, for the administration of intravitreal injection (eye)

Synergetics Inc

Sterilization certificates could not be validated by the supplier

June 12, 2024 · DeviceView details →
Class IOngoing

BAUSCH+LOMB, I-PACK Injection Kit with Wire Speculum, REF 18069W, for the administration of intravitreal injection (eye)

Synergetics Inc

Sterilization certificates could not be validated by the supplier

June 12, 2024 · DeviceView details →
Class IOngoing

BAUSCH+LOMB, I-PACK Injection Kit Custom, REF 18069, for the administration of intravitreal injection (eye)

Synergetics Inc

Sterilization certificates could not be validated by the supplier

June 12, 2024 · DeviceView details →
Class IOngoing

smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLUSELECT 10.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/100CZ; 8) BLUSELECT 6.0, CUFFED, FEN, 10/BX, Item Number 101/802/060CZ; 9) BLUSELECT 7.0, CUFFED, FEN, 10/BX, Item Number 101/802/070CZ; 10) BLUSELECT 7.5, CUFFED, FEN, 10/BX, Item Number 101/802/075CZ; 11) BLUSELECT 8.0, CUFFED, FEN, 10/BX, Item Number 101/802/080CZ; 12) BLUSELECT 8.5, CUFFED, FEN, 10/BX, Item Number 101/802/085CZ; 13) BLUSELECT 9.0, CUFFED, FEN, 10/BX, Item Number 101/802/090CZ; 14) BLUSELECT 10.0, CUFFED, FEN, 10/BX, Item Number 101/802/100CZ; 15) BLUSELECT 6.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/060CZ; 16) BLUSELECT 7.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/070CZ; 17) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075; 18) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075CZ; 19) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080; 20) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080CZ; 21) BLUSELECT 8.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/085CZ; 22) BLUSELECT 9.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/090CZ; 23) BLUSELECT 10.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/100CZ; 24) BLUSELECT 6.0, CUFFED, FEN, 1/EA, Item Number 101/812/060CZ; 25) BLUSELECT 7.0, CUFFED, FEN, 1/EA, Item Number 101/812/070CZ; 26) BLUSELECT 7.5, CUFFED, FEN, 1/EA, Item Number 101/812/075CZ; 27) BLUSELECT 8.0, CUFFED, FEN, 1/EA, Item Number 101/812/080CZ; 28) BLUSELECT 8.5, CUFFED, FEN, 1/EA, Item Number 101/812/085CZ; 29) BLUSELECT 9.0, CUFFED, FEN, 1/EA, Item Number 101/812/090CZ; 30) BLUSELECT 10.0, CUFFED, FEN, 1/EA, Item Number 101/812/100CZ; 31) BLUSELECT 6.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/060; 32) BLUSELECT 7.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/070; 33) BLUSELECT 7.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/075; 34) BLUSELECT 8.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/080; 35) BLUSELECT 8.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/085; 36) BLUSELECT 9.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/090; 37) BLUSELECT 10.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/100; 38) BLUSELECT 6.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/060; 39) BLUSELECT 7.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/070; 40) BLUSELECT 8.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/080; 41) BLUSELECT 9.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/090; 42) BLUSELECT 10.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/100; 43) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/060CZ; 44) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/070CZ; 45) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/075CZ; 46) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/080CZ; 47) BLUSELECT 8.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/085CZ; 48) BLUSELECT 9.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/090CZ; 49) BLUSELECT 10.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/100CZ; 50) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/060CZ; 51) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/070CZ; 52) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/075CZ; 53) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/0

Smiths Medical ASD Inc.

There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.

June 7, 2024 · DeviceView details →
Class IOngoing

smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect trach tube, Item Number 101/561/070; 2) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect trach tube, Item Number 101/561/080; 3) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect trach tube, Item Number 101/561/090; 4) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/070; 5) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/080; 6) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/090; 7) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/070; 8) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/080; 9) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/090; 10) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/070; 11) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/080; 12) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/090

Smiths Medical ASD Inc.

There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.

June 7, 2024 · DeviceView details →
Class IOngoing

smiths medical portex BLUgriggs Percutaneous Dilation Procedural Trays with BLUselect tracheotomy tubes, labeled as follows: 1) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 8.0mm BLUselect trach tube (drugs, US ONLY), Item Number 101/540/080; 2) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 9.0mm BLUselect trach tube (drugs, US ONLY), Item Number 101/540/090; 3) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 7.0mm BLUselect trach tube, Item Number 101/541/070; 4) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 8.0mm BLUselect trach tube, Item Number 101/541/080; 5) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 9.0mm BLUselect trach tube, Item Number 101/541/090; 6) BLUgriggs Percutaneous Dilation Procedural Kit with 7.0mm BLUselect trach tube, no forceps, Item Number 101/543/070; 7) BLUgriggs Percutaneous Dilation Procedural Kit with 8.0mm BLUselect trach tube, no forceps, Item Number 101/543/080; 8) BLUgriggs Percutaneous Dilation Procedural Kit with 9.0mm BLUselect trach tube, no forceps, Item Number 101/543/090; 9) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 7.0mm BLUselect SuctionAid trach tube, Item Number 101/891/070; 10) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 8.0mm BLUselect SuctionAid trach tube, Item Number 101/891/080; 11) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 9.0mm BLUselect SuctionAid trach tube, Item Number 101/891/090; 12) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 7.0mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/892/070; 13) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 8.0mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/892/080; 14) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 9.0mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/892/090; 15) BLUgriggs Percutaneous Dilation Procedural Kit with 7.0mm BLUselect SuctionAid trach tube, no forceps, Item Number 101/893/070; 16) BLUgriggs Percutaneous Dilation Procedural Kit with 8.0mm BLUselect SuctionAid trach tube, no forceps, Item Number 101/893/080; 17) BLUgriggs Percutaneous Dilation Procedural Kit with 9.0mm BLUselect SuctionAid trach tube, no forceps, Item Number 101/893/090

Smiths Medical ASD Inc.

There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.

June 7, 2024 · DeviceView details →
Class IOngoing

SENSE XL TORSO COIL 1.5T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567141882, 453567141883. Coils may be included in kits with Mode Numbers: 989603014351 & 989603014352.

Philips North America Llc

Potential for coils to heat up and harm patients (burn).

June 5, 2024 · DeviceView details →
Class IOngoing

SENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567394941, 453567394942, 453567394943, 453567394945, 459801593182. Coils may be included in kits with Mode Numbers: 989603050641, 989603050642.

Philips North America Llc

Potential for coils to heat up and harm patients (burn).

June 5, 2024 · DeviceView details →
Class IOngoing

SENSE XL TORSO COIL 1.5T Mk2. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567502281, 453567502282, 453567502283, 453567502284, 453567502285, 453567502286, 459801585212.

Philips North America Llc

Potential for coils to heat up and harm patients (burn).

June 5, 2024 · DeviceView details →
Class IOngoing

Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX

Smiths Medical ASD Inc.

There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.

May 31, 2024 · DeviceView details →
Class IOngoing

paraPAC plus 310 ventilator kit with internal PEEP and CPAP, Item Number P310N

Smiths Medical ASD Inc.

Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.

May 31, 2024 · DeviceView details →
Class IOngoing

Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump

Abiomed, Inc.

Nine (9) Impella CP pumps failed inspection and were inadvertently released.

May 31, 2024 · DeviceView details →
← PreviousPage 31 of 50Next →