  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 501–520 of 1,000 recalls

Class ITerminated

Draeger Jaundice Meter JM-105 The device is intended for use in hospitals, clinics or doctor s offices under a physician s supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements. The Jaundice Meter is not intended for home use. The Jaundice Meter may only be used at the sternum measurement site for Physician s office applications.

Draeger Medical Systems, Inc.

Users have misinterpreted the display for out of range measurement indicated by the blinking "-0-" to mean a zero measurement

May 15, 2018 · DeviceView details →
Class ITerminated

AirLife Resuscitation Device, Adult labeled as the following: a. Adult, with mask, oxygen reservoir bag, Part Number 2K8004; b. Adult, with mask, oxygen reservoir bag, CO2 detector, PEEP valve, Part Number 2K8035C2; c. Adult, with mask, 28" large bore tubing, Part Number 2K8017; d. Adult, with mask, 40" oxygen reservoir tubing, Part Number 2K8005; e. Adult, with mask, 40" oxygen reservoir tubing, PEEP valve, Part Number 2K8036; f. Adult, without mask, 40" oxygen reservoir tubing, Part Number 2K8001; g. Adult, with mask, oxygen reservoir bag, manometer, PEEP valve, Part Number 2K8035M; h. Adult, with mask, oxygen reservoir bag, CO2 detector, Part Number 2K8004C2; Manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own. The devices are intended for use as single patient use resuscitator devices for pulmonary resuscitation.

Vyaire Medical

An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.

May 8, 2018 · DeviceView details →
Class ITerminated

AirLife Resuscitation Device, Pediatric labeled as the following: a. Pediatric, with mask, 28" large bore tubing, pressure-relief valve, Part Number 2K8018; b. Pediatric, with mask, 40" oxygen reservoir bag, pressure-relief valve, Part Number 2K8008; c. Pediatric, Resuscitation kit with neonatal, infant and pediatric masks, Part Number RE1DK5445D; d. Pediatric, with mask, oxygen reservoir bag, pressure-relief valve, PEEP valve, Part Number 2K8039 Manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own. The devices are intended for use as single patient use resuscitator devices for pulmonary resuscitation.

Vyaire Medical

An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.

May 8, 2018 · DeviceView details →
Class ITerminated

WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.

Biosense Webster, Inc.

When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.

May 7, 2018 · DeviceView details →
Class ITerminated

Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.

CareFusion 303, Inc.

The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

April 13, 2018 · DeviceView details →
Class ITerminated

Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3 Indicated for providing short-term hemodynamic support in patients with advanced refractory left ventricular heart failure.

Abbott

Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death.

April 5, 2018 · DeviceView details →
Class ITerminated

BD Vacutainer¿ Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI: 50382903678841); 3. 367886 (UDI: 50382903678865); 4. 368040 (UDI: 50382903680400); 5. 366667 (UDI: 50382903666671); 6. 367880 (UDI: 50382903678803)

Becton Dickinson & Company

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

March 22, 2018 · DeviceView details →
Class ITerminated

BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)

Becton Dickinson & Company

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

March 22, 2018 · DeviceView details →
Class ITerminated

BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)

Becton Dickinson & Company

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

March 22, 2018 · DeviceView details →
Class ITerminated

BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)

Becton Dickinson & Company

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

March 22, 2018 · DeviceView details →
Class ITerminated

Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.

Draeger Medical, Inc.

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

March 1, 2018 · DeviceView details →
Class ITerminated

Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Draeger Medical, Inc.

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

March 1, 2018 · DeviceView details →
Class ITerminated

Fabius Tiro, anesthesia machine, catalog no. 8606000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Draeger Medical, Inc.

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

March 1, 2018 · DeviceView details →
Class ITerminated

Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Draeger Medical, Inc.

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

March 1, 2018 · DeviceView details →
Class ITerminated

MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.

Micro Therapeutics Inc, Dba Ev3 Neurovascular

There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.

February 26, 2018 · DeviceView details →
Class ITerminated

EBI Osteogen Implantable Bone Growth Stimulator The OsteoGen Implantable Bone Growth Stimulator is indicated in the treatment of long bone nonunions.

Zimmer Biomet, Inc.

Lack of adequate validation and controls to ensure that product cleanliness parameters were consistently met.

February 19, 2018 · DeviceView details →
Class ITerminated

SpF PLUS-Mini (60 ¿A/W) Implantable Spinal Fusion Stimulator SpF PLUS-Mini Implantable Fusion Stimulators are indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels

Zimmer Biomet, Inc.

Lack of adequate validation and controls to ensure that product cleanliness parameters were consistently met.

February 19, 2018 · DeviceView details →
Class ITerminated

bioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit bioMerieux VITEK 2 AST-GP75 Test Kit bioMerieux VITEK 2 AST-GP78 Test Kit Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicrobial susceptibility testing of bacteria and yeast.

bioMerieux, Inc.

Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.

January 26, 2018 · DeviceView details →
Class ITerminated

1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics

Beckman Coulter Inc.

A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.

January 12, 2018 · DeviceView details →
Class ITerminated

COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software

Beckman Coulter Inc.

A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.

January 12, 2018 · DeviceView details →