  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 421–440 of 1,000 recalls

Class IOngoing

Novum IQ Syringe Pump, Product Code 40700BAX, infusion pump

Baxter Healthcare Corporation

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

August 4, 2025 · DeviceView details →
Class IOngoing

Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH

Medline Industries, LP

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

July 31, 2025 · DeviceView details →
Class IOngoing

Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH

Medline Industries, LP

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

July 31, 2025 · DeviceView details →
Class IOngoing

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH

Medline Industries, LP

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

July 31, 2025 · DeviceView details →
Class IOngoing

Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. Medline Item Numbers 10135910RH 10135936RH

Medline Industries, LP

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

July 31, 2025 · DeviceView details →
Class IOngoing

Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH

Medline Industries, LP

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

July 31, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN FXD Curve Access System Dbl, US, Material Number (UPN) M635TU80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System DBL, US, Material Number (UPN) M635TU70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN FXD Curve Access Sys Dbl, OUS, Material Number (UPN) M635TS80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSteer Access System, US, Material Number (UPN) M635TU90050;intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System SGL, OUS, Material Number (UPN) M635TS70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN FXD Curve Access System Sgl, US, Material Number (UPN) M635TU80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System ANT, US, Material Number (UPN) M635TU70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Boston Scientific Corporation

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

July 29, 2025 · DeviceView details →
Class IOngoing

Medline medical convenience kits labeled as: 1. ADULT INTUBATION TRAY, Kit Number ACC010369B; 2. ADULT ICU INTUBATION TRAY, Kit Number ACC010379A; 3. INTUBATION TRAY, Kit Number ACC010392; 4. INTUBATION KIT, Kit Number ACC010467A; 5. ADULT AIRWAY TRAY, Kit Number ACC010469; 6. TRAY 2 INTUBATION DRAWER, Kit Number ACC010480; 7. TRAY 2 INTUBATION DRAWER, Kit Number ACC010480A; 8. ADULT INTUBATION, Kit Number ACC010540; 9. DIFFICULT AIRWAY TRAY #1, Kit Number ACC010593A; 10. DIFFICULT AIRWAY TRAY #1, Kit Number ACC010593B; 11. INTUBATION TRAY, Kit Number DYNDA1847A; 12. INTUBATION TRAY (ADULT), Kit Number DYNJAA245A.

MEDLINE INDUSTRIES, LP - Northfield

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.

July 25, 2025 · DeviceView details →
Class IOngoing

Medline medical convenience kits labeled as: 1. NEWBORN RESUSCITATION KIT, Kit Number ACC010381C; 2. DRAWER 5 - POSITION 3, Kit Number ACC010532; 3. DRAWER #4C ADULT, Kit Number ACC010671; 4. DRAWER #2, Kit Number ACC010728.

MEDLINE INDUSTRIES, LP - Northfield

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.

July 25, 2025 · DeviceView details →
Class IOngoing

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System

Dexcom, Inc.

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

July 24, 2025 · DeviceView details →