  1. FDA Recalls
  2. ›Device Recalls

FDA Device Recalls

Recalls of medical devices, diagnostic equipment, implants, and surgical instruments.

Classification
AllClass IClass IIClass III
Status
AllOngoingCompletedTerminated
Date
All timePast 30 daysPast 12 months

Showing 341–360 of 1,000 recalls

Class ITerminated

CARESCAPE Respiratory Module E-sCO

GE Healthcare, LLC

Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

September 27, 2019 · DeviceView details →
Class ITerminated

CARESCAPE Respiratory Module E-sCOVX

GE Healthcare, LLC

Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

September 27, 2019 · DeviceView details →
Class ITerminated

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

Philips Medical Systems (Cleveland) Inc

An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.

September 19, 2019 · DeviceView details →
Class ITerminated

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

Philips Medical Systems (Cleveland) Inc

An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.

September 19, 2019 · DeviceView details →
Class ITerminated

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

Philips Medical Systems (Cleveland) Inc

An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.

September 19, 2019 · DeviceView details →
Class ITerminated

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

Philips Medical Systems (Cleveland) Inc

An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.

September 19, 2019 · DeviceView details →
Class ITerminated

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.

Philips Medical Systems (Cleveland) Inc

An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.

September 19, 2019 · DeviceView details →
Class ITerminated

BodyGuard Microset REF A120-003XYVA, Sterile EO, Tubing: 150 in. (380 cm), Prime Vol: approx. 8ml Product Usage: The BodyGuard Infusion Pump System is designed to transfer medication and fluids intravenously.

CME America, LLC

When using a specific tubing set with infusion pump system, may result in under deliver of fluids.

September 16, 2019 · DeviceView details →
Class ITerminated

Rosa Brain 3.0, Model No. ROSA 3.0. The device is delivered assembled in a wooden case and protected with plastic bubble and plastic film. Seals are put on the crate. The product is not shipped sterile. The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons.

MEDTECH SAS

The instrument holder may be sent on a trajectory that is not within the intended target. If it is not corrected, the associated device may be placed incorrectly.

September 10, 2019 · DeviceView details →
Class ITerminated

VNS Therapy, SenTiva, Model # 1000, (01)05425025750405, Rx Only for vagal nerve stimulation (VNS) therapy

LivaNova USA Inc

Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.

August 22, 2019 · DeviceView details →
Class ITerminated

CentriMag Acute Circulatory Support System Motor, U.S. Model 102956 and OUS Model 201-10002, a component of the CentriMag Extracorporeal Blood Pumping System.

Thoratec Switzerland GMBH

Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).

August 22, 2019 · DeviceView details →
Class ITerminated

Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS License and O-arm Imaging System utilizing the auto-registration (fiducial- less) workflow used in combination during a DBS (deep brain stimulation) procedure. Neurological stereotaxic Instrument.

Medtronic Navigation, Inc.

Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).

August 9, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0 mm, REF 5-22518 h) 6.0 mm, REF 5-22512 i) 6.5 mm, REF 5-22513 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm, REF 5-10316 h) 8.5 mm, REF 5-10317 i) 9.0 mm, REF 5-10318 j) 10.0 mm, REF 5-10320 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10210 f) 5.5 mm, REF 5-10111 g) 6.0 mm, REF 5-10112, REF 5-10212 h) 6.5 mm, REF 5-10113, REF 5-10213 I) 7.0 mm, REF 5-10114, REF 5-10214 j) 7.5 mm, REF V5-10115, REF 5-10115, REF 5-10215 k) 8.0 mm, REF 5-10116, REF 5-10216, REF V5-10116 l) 8.5 mm, REF 5-10117, REF 5-10217 m) 9.0 mm, REF 5-10118, REF5-10218 n) 9.5 mm, REF 5-10119 o) 10.0 mm, REF 5-10120, REF 5-10220 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009, REF 5-22109, REF 5-22209 e) 5.0 mm, REF 5-22010, REF 5-22110, REF 5-22210 f) 5.5 mm, REF 5-22011, REF 5-22111, REF 5-22211 g) 6.0 mm, REF 5-22212, REF 5-22312, REF 5-22112 h) 6.5 mm, REF 5-22313, REF 5-22213, REF 5-22013, REF 5-22113 i) 7.0 mm, REF 5-22214, REF 5-22314, REF 5-22014, REF 5-22114 j) 7.5 mm, REF 5-22215, REF 5-22315 k) 8.0 mm, REF 5-22216, REF 5-22316 l) 8.5 mm, REF 5-22217 m) 9.0 mm, REF 5-22218 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g) 5.0 mm, REF 5-10410 h) 5.5 mm, REF 5-10411 i) 6.0 mm, REF 5-10412 j) 6.5 mm, REF 5-10413 k) 7.0 mm, REF 5-10414 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
Class ITerminated

Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management

Teleflex Medical

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

July 30, 2019 · DeviceView details →
← PreviousPage 18 of 50Next →