Verathon, Inc.
When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.
Arrow International Inc
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.